Review Team (February 2016)
BVit-03 (February 2016) FOR ERP USE ONLY DO NOT DISTRIBUTE
AOAC- SPIFAN-2012-VITAMINS
Validation Parameter
Limit of Quantification
Test Group Name
Date
TUV/BLR/F-360 2015\January\Vitamins\01.01.2015 01.01.2015 Prasad BKVR/Kesav Kumar N Manjunath K/Prasad BKVR TUV/BLR/F-99
B-Complex vitamins Thiamine Aqueous
Test Name
Data Path File
Matrix
Raw-Data File Ref. CRM Data File Ref.
Limit
Equipment Test Method
Analyst
LC-MS/MS 6460A (BLR/PS-CRS/006.03)
Operator
Repeatability at LOQ 5 ug/l standard Sample Blank No.
Mean Retention Time (min)
m/z
Signal (ug/l)
Caluclation
265.2>122.2 265.2>122.2 265.2>122.2 265.2>122.2 265.2>122.2 265.2>122.2 265.2>122.2 265.2>122.2 265.2>122.2 265.2>122.2
1.823 1.823 1.823 1.823 1.823 1.823 1.823 1.823 1.823 1.823
4.608 4.785 4.453 4.814 4.194 4.135 4.675 4.774 4.099 3.911
1 2 3 4 5 6 7 8 9
Mean (μg/l)
4.44
SD
0.3339
LOD (Mean +3 sigma) LOQ (Mean + 10sigma)
5.45 7.78 7.51
RSD %
5
Standard Conc. (μg/l)
88.9
Accuracy %
Accuracy = 70 - 125 %
Criteria
10
RSD% = <15%
Page 21