Review Team (February 2016)
BVit-03 (February 2016) FOR ERP USE ONLY DO NOT DISTRIBUTE
AOAC- SPIFAN-2012-VITAMINS
Validation Parameter
Limit of Quantification
Test Group Name
Date
01.01.2015 2015\January\Vitamins\01.01.2015 Prasad BKVR/Kesav Kumar N TUV/BLR/F-360 TUV/BLR/F-99 Manjunath K/Prasad BKVR
B-Complex vitamins Niacinamide Aqueous
Test Name
Data Path File
Matrix
Raw-Data File Ref. CRM Data File Ref.
Limit
Equipment Test Method
Analyst
LC-MS/MS 6460A (BLR/PS-CRS/006.03)
Operator
Repeatability at LOQ 5 ug/l standard Sample Blank No.
Mean Retention Time (min)
m/z
Signal (ug/l)
Caluclation
123.2>80.3 123.2>80.3 123.2>80.3 123.2>80.3 123.2>80.3 123.2>80.3 123.2>80.3 123.2>80.3 123.2>80.3 123.2>80.3
2.315 2.315 2.315 2.315 2.315 2.315 2.315 2.315 2.315 2.315
4.949 4.916 4.539 4.639 4.896 4.580 4.966 4.491 4.725 4.555
1 2 3 4 5 6 7 8 9
Mean (μg/l)
4.73
SD
0.1886
LOD (Mean +3 sigma) LOQ (Mean + 10sigma)
5.29 6.61 3.99
RSD %
5
Standard Conc. (μg/l)
94.5
Accuracy %
Accuracy = 70 - 125 %
Criteria
10
RSD% = <15%
Page 23