SPIFAN Nutrients ERP Book_9-29-15
FOS-04 FOR ERP USE ONLY DO NOT DISTRIBUTE
Table 2-4 Method comparison
997.08 1
999.03 2
New method
Reference 3
Number of sample preparations and analyses needed to complete testing
1
3
3
1
Instrument Run time (min)
45
83x3=249
NAP
65
Can test samples with sucrose:fructan ratio >3:1 Requires post-column reagent addition
Yes
No
Yes
Yes
No
No
NAP
Yes
Applicable to FOS, oligofructose, and inulin
Yes
Yes*
Yes*
Yes
Requires extraction step
No No No
Yes
Yes
Yes Yes Yes
Requires SPE
No No
No No
Requires dry-down
Requires knowledge of commodity type for accurate results**
Yes
Yes
Yes
Yes
Method bias for any commodities
No
No
Yes
No
Uses borate trap
Yes
No
No
No
0.03% (90.8-115% recovery)
0.014% (83-101% recovery)
LOQ (% on RTF basis)
0.5%
0.5%
%RSD (Intermediate Precision?)
2.04-7.12%
5.79%
1.25%
4.22-8.37%
1 Stated method performance taken from Covance labs website effective 10/15/2010. 2 Stated method performance taken from Covance labs website effective 01/21/2005.
*Method accuracy compromised by usage of set DP factors of 4 and 10 (actual range is from 3.7-30). **Limits utility at a regulatory level where the type of material may be unknown or testing for off-label adulteration may be desired. The new method circumvents this via the qualitative ID methodology in the Appendix. Reference #3 is pursuing an alternate method to circumvent this problem.
Conceptually this methodology is the similar to that of Cuany, et. al. (3) However several time savings are gained from the elimination of extraction, SPE, and sample dry-down. An additional time savings is generated in a shorter instrument cycle time (45 minutes versus 65 minutes). Some of this is due to the incorporation of a borate trap to the analysis in place of a standard PA1 guard. Borate negatively impacts HPAEC-PAD chromatography, creating peak distortion and other issues previously observed (3, 5). Furthermore, the addition of the internal standard (as suggested in 997.08) allows for a non-volumetric dilution and a subsequent improvement in LOQ.
1. Linearity – For the 43 standard curves examined no calibration errors greater than + 5% were observed (see figure 2-3). As expected the largest errors were observed in the lowest level standards. However the pattern of errors does not indicate any systematic trends.
Page 11 of 14
Made with FlippingBook